For operational installations a regular maintenance on a 6- or 12-monthly interval is recommended depending on the relevance of the derived data. The core maintenance should include a visual inspection of all cables and connectors, mechanic fixtures and of the absorbing foam used inside the acoustic shield. Further each loudspeaker should be checked in view of efficiency and phase stability. Test signals with given spectral characteristics should be entered to the electronics to check the internal electronic noise and the signal analysis. A comprehensive check of the measured data in view of plausibility for a period of several weeks is mandatory to identify occasionally occurring operational constraints.