The sensor head heating consists of two different heating devices, a direct heating of each transducer and a heating of the sensor head structure:
A) The transducers are heated by PTC elements which draw more power at lower PTC temperature. As a consequence, the PTC heating power is maximum immediately after the heating or the complete sensor have been switched on. With increasing time, the heating power will fall to a terminal value according to the outside temperature.
B) The sensor head structure is heated by a constant value by inside heating wires.

If operated outside, the surfaces of the transducers and of the sensor head will often show only a minimum excess temperature as compared to the ambient temperature due to the surface cooling.

If a uSonic is started and the sensor head heating is set to the parameter value ‘2’ (automatic) or ‘3’ (automatic with data availability check) and no valid measured values are available, the sensor head heating switches on automatically. In this case, the sensor head heating control assumes that the sensor cannot provide any data due to icing.

Attention: If a uSonic is operated in the laboratory or inside the shipping carton and no valid measured values are available (e.g. because measuring paths are blocked or because there are strong reflections in the shipping carton), the automatic control mentioned above can switch on the sensor head heating.