The accuracy of sonics for the wind measurement is determined mainly by shadow effects at the transducers and flow distortion effects at the sensor head structures. All measuring paths are influenced by these effects corresponding to their position and alignment relative to the inflow angle. Beside of this, also the so called zero-calibration must be regarded which compensates a possible offset in the wind components.

For conventional uSonics, wind tunnel derived corrections schemes are applied to compensate these shadow and flow distortion effects. The typically achieved accuracy is 2 % for wind speeds >5 m/s, or 0.1 m/s for vertical inflow angles within ± 15°.

For multi path uSonics (uSonic-3 Class A MP, uSonic-3 Cage MP) the most advantageously positioned paths are automatically determined and selected. The typically achieved accuracy is 1 % for wind speeds >5 m/s, or 0.05 m/s for vertical inflow angles within ± 25°.

It should be noted that the given accuracies apply to perfectly zero-calibrated sensors.